If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual EUTECH INSTRUMENTS FIELD PHORPCONDUCTIVITY MONITORS. We hope that this EUTECH INSTRUMENTS FIELD PHORPCONDUCTIVITY MONITORS user guide will be useful to you.
Manual abstract: user guide EUTECH INSTRUMENTS FIELD PHORPCONDUCTIVITY MONITORS
Detailed instructions for use are in the User's Guide.
[. . . ] Press the ENTER key to confirm the calibration while in calibration mode.
pH FM1 Continuous Monitor
CAL
CAL
ENTER
ENTER
2. 2
CAL
Keys for ORP Monitor
The CAL key switches monitor into calibration mode from measurement mode. See page 5 for details.
CAL ENTER
ENTER
Press the ENTER key to confirm the calibration while in calibration mode. The Down/Decrement key scrolls value down in calibration mode.
ORP FM1 Continuous Monitor
2. 3
CAL
Keys for Conductivity Monitor
The CAL key switches monitor into calibration mode from measurement mode. In the calibration mode, pushing CAL key confirms the calibration value and returns to measurement mode. [. . . ] Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measurement. If the electrode has been stored dry, wet the probe in tap water for 10 minutes before calibrating or taking readings to saturate the ORP electrode surface and minimize drift. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. IMPORTANT: ORP offset adjustment is limited to ±100 mV from the original reading. NOTE: All new calibration data will over-ride existing calibration data.
5
5. 2
ORP Calibration
1. The monitor will Immerse ORP electrode into a calibration standard solution, be in ORP measurement mode.
for example, 255 mV
2. Do not wipe the probe; this causes a build-up of electro-static charge on the surface. "CA" will flash on the display for a second and then a value close to the calibration standard will flash. Press the or keys to adjust the display to match the calibration standard value. The reading should now display the same value as the calibration standard. See Figure 3.
Press
Press
Press
Figure 5 - ORP Calibration Procedure
6
6
CONDUCTIVITY CALIBRATION
This section is applicable for the following models:
Description Conductivity FMCON 1 with Conductivity electrode and 110/120 VAC power adapter. "CA" will flash on the display for a second and then a value close to the calibration standard will flash. Press the or keys to adjust the display to match the calibration standard value. The reading should now display the same value as the calibration standard. See Figure 4.
into a standard solution, for example, 4 mS
Press
Press
Press
Figure 7 - Conductivity Calibration Procedure
8
7
MONITORING
Once the monitor is calibrated it is ready for monitoring. See figure below.
Mounting Eyelets
CAL
ENTER
pH FM1 Continuous Monitor
8
ELECTRODE CARE AND MAINTENANCE
Since your pH/ORP electrode is susceptible to dirt and contamination, clean it every one to three months depending on the extent and condition of use. Depending on the environment and the solution it is immersed in, it may last from 6 to 12 months, or longer. Knowing when to change your pH/ORP electrode usually comes after some experience. It is best to have another pH/ORP tester such as pHScan/ pHTestr 1, Waterproof pHScan/ pHTestr WP2, or ORPScan/ ORPTestr as reference. 8. 1 pH/ORP electrode storage
For best results, always keep the pH/ORP bulb wet. Use the protective electrode storage bottle filled with electrode storage solution (0. 1M KCl) or pH 7 for overnight and long term storage. If left dry, soak in tap water or pH 7. 0 buffer solution overnight.
9
8. 2 · ·
pH/ORP electrode cleaning Salt deposits: Dissolve the deposits by immersing the electrode in tap water for ten to fifteen minutes. [. . . ] 12. 88 mS KCl Calibration Solution in 480-ml leakproof bottle (1 pint) 2, 764 µS KCl Calibration Solution in 480-ml leakproof bottle (1 pint) 10 µS conductivity standard sachet, 20 ml x 20 pcs 447 µS Conductivity Sachets (20 units x 20 ml per box) 1, 413 µS Conductivity Sachets(20 units x 20 ml per box) 2, 764 µS Conductivity Sachets(20 units x 20 ml per box) 15, 000 µS Conductivity Sachets(20 units x 20 ml per box) Eutech Instruments Ordering Code EC-BU-4BT EC-BU-7BT EC-BU-10BT EC-BU-4BS EC-BU-7BS EC-BU-10BS EC-CON-1288BT EC-CON-2764BT EC-CON-10BS EC-CON-447BS EC-CON-1413BS EC-CON-2764BS EC-CON-15000BS Oakton Instruments Ordering Code 00654-00 00654-04 00654-08 35653-01 35653-02 35653-03 00606-10 00653-20 35653-09 35653-10 35653-11 35653-12 35653-13
14
13 WARRANTY
The Field Monitor series is supplied with a 1-year warranty from manufacturing defects and electrodes for 6 months from the date of purchase unless otherwise stated. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return freight pre-paid and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse. Exclusions The warranty on your instrument shall not apply to defects resulting from: · · · Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products
14 RETURN OF ITEMS
Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS FIELD PHORPCONDUCTIVITY MONITORS
Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual EUTECH INSTRUMENTS FIELD PHORPCONDUCTIVITY MONITORS will begin.