Detailed instructions for use are in the User's Guide.
[. . . ] Alpha DO 2000 PPG
Processor/Transmitter Dissolved Oxygen ( ppm / m/L / % )
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTOCAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. [. . . ] Press ENTER key to confirm your selection NOTE: Accepting the zero point will reset the zero offset settings made in OFS menu.
8. The temperature compensated sensor current (in nA) for the second point will be displayed. Once the reading stabilizes, the flashing symbols disappear and "READY" comes on. Press or key to select the desired saturation ("SAT %") or concentration ("Con ppm" or "Con mg/l"). NOTES: · · If you entered the Calibration mode from the Setup mode, the transmitter will return to the setup menu. If there is a calibration error the transmitter displays "ERR". In this case, press and keys simultaneously (escape) to restart the calibration beginning from step 1. When calibrating with manual temperature compensation, the transmitter automatically changes from the preset process temperature to the calibration temperature. After leaving the Calibration mode, the transmitter switches back to the process temperature (for setting the calibration temperature and the process temperature, see section 6. 3).
·
17
Instruction manual
Alpha DO 2000
5. 3 View Probe Condition
The probe condition is an indication of its current working condition with respect to its original condition. The probe condition is calculated and stored in the controller each time calibration is done. You can view probe condition without entering into calibration mode.
1
%
MEAS
89. 5
CAL
24. 5 ATC
2
000
C. CD
ENT CAL
3
CAL
DO
ENT CAL
ENT
4
ENT
CAL
5
ENT
CAL
6
%
0. 0
95. 5
100. 0
ADJ
1. NOTE: If no key is pressed for 30 seconds during the above steps, the controller automatically returns to the Measurement mode.
18
Instruction manual
Alpha DO 2000
6. Setup Mode
6. 1 Enter Setup Mode
In the Setup mode the transmitter can be configured to your individual requirements.
1. Set the security code with or key to: - "SCD 22" if you want to change parameter settings
%
°C
MEAS
13. 0
ENT
- "SCD 000" if you only want to view parameter settings (view only mode) 3. Press ENT key.
ENT
SETUP HOLD
OFS
ENT
Offset settings see section 6. 2 Temperature settings see section 6. 3 Relay A (set point 1) settings see section 6. 4 Relay B (set point 2) settings see section 6. 4 Controller settings see section 6. 5
SETUP HOLD
ENT
SETUP HOLD
ENT
SETUP HOLD
ENT
SETUP HOLD
ENT
SE TUP
ENT
HOLD
Current output 1 settings see section 6. 6 Current output 2 settings see section 6. 7
SETUP HOLD
ENT
SETUP HOLD
ENT
Wash relay settings see section 6. 8
SETUP HOLD
ENT
Configuration settings see section 6. 9
SETUP HOLD
ENT
DO
CAL
Calibration see section 6. 10
NOTE: To exit Setup mode at any time press and keys simultaneously (escape). The transmitter returns to the Measurement mode.
19
Instruction manual
Alpha DO 2000
6. 2 Electrode Offset Settings (OFS)
The electrode offset settings allow you to zero the transmitter without a sensor. For real zero calibration, use the calibration procedure as described in section 5. If no simulator is used, a 22 k NTC must be connected to pins 24 and 25 of the terminal connector on the backside of the unit (see section 3. 2).
1 2 5 6
ENT
ppt
SETUP
HOLD
OFS
ENT
ENT
SETUP
HOLD
NO
OFS
2
ENT
SETUP
HOLD
SETUP
HOLD
00 .
SAL
BAR
PRES
6
ENT
SETUP
HOLD
7
1. 013
BAR
ENT
SETUP
HOLD
YES
OFS
ENT
This screen will not be visible if DO Unit of Measurement is set to % (In Configuration settings)
SETUP
HOLD
H9
PRES
6
SETUP
HOLD
3
00 .
OFS
ENT
%
SETUP
HOLD
PSI
PRE S
SETUP
HOLD
4
00 .
ADJ
%
ENT
1. From the setup mode select the "OFS" screen and then press ENT key. Selecting to modify offset value: By default, controller shows "NO OFS". Press or key to select "NO" or "YES" for offset adjustment. - Select "NO OFS" to skip offset adjustment and move to the next step. [. . . ] The warranty card must be endorsed by the Authorized Distributor at the point of sale. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return freight pre-paid and correction will be made without charge. Thermo Scientific/ Eutech will determine if the product problem is due to deviations or customer misuse. Exclusions The warranty on your instrument shall not apply to defects resulting from: · Improper or inadequate maintenance by customer · Unauthorized modification or misuse · Operation outside of the environment specifications of the products
10. 2 Packaging / Scope of delivery
The instrument is packaged in a corrugated box with an instruction manual and the following accessories: Wall mount version: · Plug, nylon, red, 2 pcs · PG13. 5, cable gland, 3 pcs · Connectors, one each of 3 way (5. 08 mm), 9 way (5. 08 mm), 13 way (3. 5mm), 8 way (3. 5mm). [. . . ]