User manual EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER. We hope that this EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER user guide will be useful to you.
Manual abstract: user guide EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER
Detailed instructions for use are in the User's Guide.
[. . . ] EUTECH INSTRUMENTS
3
Diagram and Dimensions (in mm)
2-Wire Temperature Transmitter
Operating Instructions
4
Installation
Operating with one or several Transmitters:
Figure 1 - Operating with 1 Transmitter
1
Preface
Thank you for purchasing the Eutech 2-wire Temperature Transmitter. [. . . ] · · · · · EC-PH-XTR; EC-ORP-XTR; EC-CON-XTR; EC-DO-XTR; and EC-TEMP-XTR.
The information presented in this manual is subject to change without prior notice, as improvements are made to the product. While Eutech will endeavor to try its best to inform the market, this does not represent a commitment on the part of Eutech Instruments Pte Ltd. Eutech Instruments does not accept any responsibility for damage or malfunction of the unit due to improper installation or operation of the Transmitter. Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail: marketing@eutechinst. com; Home page: http://www. eutechinst. com
Figure 2 - Operating with several transmitters
2
Introduction
5
Electrical Connections
The Transmitter is used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen or Temperature, in conjunction with an appropriate sensor. The Transmitter is housed in a IP65 enclosure, with openings for input and 4-20mA output.
68X256904 Rev 1 10/99
6
Before Calibration
Remove cover of Transmitter and ensure that electrode is correctly connected as above. Connect Amp meter to the 4/20mA output of Transmitter (Appendix 1).
7
1) 2) 3) 4) 5) 6) 7) 8) 9)
Temperature Transmitter Calibration
Immerse Temperature probe in water bath at varying temperatures and observe Amp meter display; Amp meter should read corresponding mA output as follows: In 0oC, the output is 4 mA; In 25oC, the output is 8 mA; In 50oC, the output is 12 mA; In 80oC, the output is 16. 8 mA; In 100oC, the output is 20 mA; To calibrate, locate "Temp adjustment" trimmer; Using a fine screwdriver, slowly turn trimmer till Amp meter reads correct current output;
pH Transmitter Combination pH Electrode, General, integral 5m cable Combination pH Electrode, w/PMP, integral 5m cable
Figure 4 - Method Two
10
pH
Ordering Information
Item Cat Code ECPHXTR ECARTSO05B ECARGTSO05B
NB: For other Temperature values, expected current output is based on: mA = {(16/100) x Temperature value} + 4mA
Dissolved Oxygen
Item Dissolved Oxygen Transmitter General Purpose D. O. [. . . ] probe, 0 10 ppm, Industrial, integral 5m cable
ORP
Model Range Accuracy Calibration Input Output Operating Voltage Load
Item ORP Transmitter Combination Platinum integral 5m cable ORP Electrode, w/PMP, Cat Code ECORPXTR ECHTPTTSO05B ECHTAUTTSO05B
Combination Gold ORP Electrode, w/PMP, integral 5m cable
Conductivity
Item Conductivity Transmitter 2-Pin SS, k=1. 0, integral 3m cable Cat Code ECCONXTR ECCONSEN46
Housing
77 x 77 x 28 mm, field mountable
Temperature
Item Cat Code ECTEMPXTR ECPT10001M or ECPT10005M
9
Appendix 1
Temperature Transmitter Pt 100 probe, SS316
Connecting Amp Meter for Calibration
Power Supply
Item 120V Power Supply 220V Power Supply Cat Code EC-120-XTR EC-220-XTR
Figure 3 - Method One
68X256904 Rev 1 10/99 [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER
Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual EUTECH INSTRUMENTS 2-WIRE TEMPERATURE TRANSMITTER will begin.